SKX-2000G ECG simulator use
The simulator can produce the following waveforms, the first waveform as a code
representing
1, the normal ECG waveform
2, positive and negative triangular waveform
3, arrhythmia triangular waveform simulation
4, square wave, square wave measurement using scanning speed
5, sine wave, amplitude-frequency characteristic measurement
6, simulated respiratory waveform
7, simulated fetal heartbeat (another optional tooling)
8, to set the size of the signal amplitude
9 with pacing ECG signal
10, can change the width of the bottom of the triangular wave
11, can change the T-wave amplitude, heart rate, QRS wave amplitude, QRS wave
of the ECG waveform width
12, pacing signal
13, a triangular wave superimposed sinusoidal 2Hz
Note:
1, the simulator automatically generated after power waveform a normal ECG
waveform.
2, the simulator LED display tube, to prevent users forget to turn off the
power during use, system design within 24 hours of when the button is not
operated, CPU will enter standby for power saving, re-boot will work. Press any
button and the time to zero.
3, ECG connection: RA-R (right), LA-L (left), LL-F (left leg), RL-RF (right
leg), C1-C6 chest leads
4, the monitor connection: RA-hand (white), LA-left (black), LL-left (red),
RL-foot (green), C1-C6 chest leads (brown)
5, three lead connection method: RA-hand (white), LA-left (black), LL-left
(red)
6, corresponding to European standard connection method: L-LA R-RA RF (N)-RL
F-LL CV
Key Description
A total of four buttons, followed by selection keys to increase the key, down
key, the Enter key, in addition to a set of key combinations
Select button:
This button is used to select the parameter to be changed, a total of four LED
tubes to display the four codes represent the content displayed
A code representing the waveform ,2-4 represents a parameter to change (2 is
the value one hundred, three on behalf of ten, four representatives bits)
LED tube right lower leg of the highlights, which means that now the selected
content; can be changed.
Increase key:
When using the Select button to select a good to change the content, use this
key parameter changes.
Decrease key:
When using the Select button to select a good to change the content, use this
key parameter changes.
Enter key:
When the parameter change is completed, this will confirm button to confirm
changes to the parameters and generate the corresponding waveform.
Key combinations:
Press the Select key, and then press the Enter key, release the Enter key,
release the Select key, then select a key combination represents, press the
Enter key will display different content.
Instructions say to use the various waveforms, respectively, the method of
operation:
A normal ECG waveform:
★ heart rate setting range
:10-200bpm; (initial value: 60 bpm)
★ signal amplitude fixed.
2, positive and negative triangular waveforms:
★ Frequency range :10-300bpm;
(initial value: 75 bpm)
★ amplitude range :0.1-4Mv (10:0.1
mV, 400:4 mV);
★ 2 modes, Mode 1: positive waveform
Mode 2: Negative waveform; through a combination of keys.
3, arrhythmia waveforms:
★ Frequency range :20-250bpm;
(initial value: 80 bpm)
★ amplitude range :0.1-4Mv (10:0.1
mV, 400:4 mV);
★ 2 modes, Mode 1: the next big wave,
Mode 2: Big wave on top. Through a combination of keys.
4, square wave:
★ Frequency range: 0.1Hz-10Hz (10:1
Hz 100:10 Hz); (Initial value: 1 Hz)
★ amplitude range :0.1-4Mv (10:0.1
mV, 400:4 mV);
5, sine wave:
★ Frequency range :1-100Hz; (initial
value: 10 Hz)
★ amplitude range :0.1-4Mv (10:0.1
mV, 400:4 mV);
★ superimposed on the waveform DC
component, the range is 0.1mV-3mV (10:0.1 mV, 400:4 mV);
★ Through the combination of keys,
select the need to change the parameters, Mode 1: Frequency; Mode 2: DC
component superimposed
(10:0.1 mV, 400:4 mV);
6, simulated respiratory waveform (on some models may no signal):
★ Frequency range :10-100 beats /
min. (Initial value: 10 bpm)
★ Please note that at this time
ignored ECG, respiration leads to RA-LL, please use the ECG without resistance
line.
7, simulated fetal heart rate (extensions)
★ Frequency range :30-240bpm;
(initial value: 120 bpm)
8, the signal amplitude settings:
★ amplitude range: 0.1mV-4mV (10:0.1
mV, 400:4 mV); (initial value: 100)
★ Please note that this amplitude
settings will affect all waveforms.
9 with pacing ECG signal
★ 2 operating modes: Mode 1: Forward
pacing pulses with ECG; Mode 2: Negative pacing pulses with ECG; through a
combination of keys.
★ pulse waveform width: 1ms-300ms;
(initial value: 60)
★ Frequency range
:20-250bpm; (initial value: 80ms)
★ pulse waveform width adjustment,
the impact of the pulse waveform in both modes.
★ pulse waveform amplitude of the
signal amplitude is set by the eight affected.
10, width adjustable triangular waveform
★ waveform amplitude: 0.1mV-4mV
(10:0.1 mV, 400:4 mV);
★ Frequency range: Fixed 60 / S
★ width adjustment range: 2ms-400ms;
(initial value: 100ms)
11, the analog ECG waveform
★ This waveform There are four
operating modes, Mode 1: You can change the T-wave amplitude; Mode 2: Set the
waveform frequency; Mode 3: Set QRS waveform amplitude; Mode 4: Set the width
of the QRS complex, between operating modes sequentially through key
combination to convert.
★ T wave amplitude setting range:
0.1mV-2mV; (initial value: 10)
★ waveform frequency setting range
:20-300bpm; (initial value: 75)
★ QRS waveform amplitude range:
0.1mV-2mV; (initial value: 1mV)
★ QRS waveform width range:
10ms-150ms; (initial value: 80ms)
12, pacing signal (pulse waveform)
★ This waveform of two kinds, namely
positive and negative directions of pulse waveform
★ units of 0.1 ms
★ amplitude size needs to be set in
the signal amplitude (8) set the amplitude size
13, a triangular wave superimposed sinusoidal 2Hz
★ triangular waveform amplitude can be
set
- 供应便携式无创血压模拟仪H2-5000系列无创血压信号模拟器
- 供应厂家直销SKX-1000F血氧模拟仪 厂家直销SKX-1000F血氧仪
- 供应胎心模拟仪可以调节强度的模拟器
- 供应胎心模拟仪/胎心信号模拟器
- 供应脑电信号肌电信号神经元信号模拟仪 SKX-8000 微弱电生理信号
- SKX-8000 微弱电信号模拟 微弱电信号模拟仪脑电模拟 微弱电信号模拟仪简化版
- SKX-2000D+患者模拟仪心电信号发生器
- SKX-2001心电检测仪监护检定仪心电监护检定 铭昇SKX-2001心电检测仪
- 供应厂家直销SKX-8000B肌电信号模拟仪 厂销SKX-8000B肌电信 SKX8000B肌电信号发生器
- 心电信号发生器患者模拟仪 心电信号发生器心电模拟仪 仪 增强型